Benchtop Calibrator
The Martel 2001 Bench Calibrator is designed to offer laboratory
grade accuracy at a hand-held calibrator price. This new unit incorporates
multiple temperature and voltage ranges and a milliamp range. The
M2001 has the ability to store up to 9 setpoints for each output range.
the setpoints can be selected manually or automatically stepped at timed
intervals. The M2001 also features a user interface similar to a Fluke 5500
calibrator as well as computer control through an RS-232 or optional IEEE 488
interface for automated production testing.
Features:
• Source or Measure up to 10 T/C or
10 RTD's with an Accuracy of up to 0.025 C
• Measure Pressure using External Fluke 700 Pressure Modules
• RTD Simulation Works with all Pulsed Transmitters
• Source DC Volts from mV to 100
VDC and Source Current to 100mA
Technical Specifications:
| Power Requirements
|
120/240 VAC
|
| Remote Interface
|
RS-232 and GPIB (IEEE 488-2 optional)
|
| Circuit Protection
|
Internal Self Resetting Fuses
|
| Display
|
16 x 2 Large Character
Alphanumeric with LED Backlight
|
| Size
|
Approx. 11.5" x 4.7" x 12"
|
| Weight
|
Approx. 5 pounds
|
| Function
|
Output Range
|
Input Range
|
Resolution
|
Accuracy**(1 year)
|
|
100mV
|
0 to 100.000mV
|
N/A
|
1uV
|
+/-0.003% +/-3 counts
|
|
1V
|
0 to 1.00000V
|
N/A
|
10uV
|
+/-0.003% +/-2 counts
|
|
10V
|
0 to 10.0000V
|
N/A
|
100uV
|
+/-0.003% +/-2 counts
|
|
100V
|
0 to 100.000V
|
N/A
|
1mV
|
+/-0.003% +/-2 counts
|
|
100mA
|
0 to 100.000mA
|
N/A
|
1uA
|
+/-0.01% +/-2 counts
|
|
T/C
|
J,K,T,E,R
S,N,B,L,U,mV
|
J,T,E,R,
S,N,B,L,U,mV
|
Out 0.1C/F
In 0.01 C/F
|
Type J Typical 0.16C
Type J Typical 0.16C
|
|
RTD*
|
PT385 (100, 200, 500, 1000)
PT392, PT3916(JIS)
Ni120, Cu10, YSI400
|
PT385 (100, 200, 500, 1000)
PT392, PT3916(JIS)
Ni120, Cu10, YSI400, 25 Ohm SPRT
|
Typical PT385-100
Out 0.1 C/F
In 0.01 C/F
|
Typical PT385-100
+/- 0.10C
+/-0.020C
|
|
Ohms
|
N/A
|
0 to 100.000
|
0.001
|
40 PPM ± 0.02
|
|
0 to 400.00
|
0 to 400.00
|
Out 0.01
In 0.001
|
+/-0.03
40 PPM +/-0.002
|
|
0 to 4000.0
|
0 to 4000.00
|
Out 0.1
In 0.01
|
+/-0.3
40 PPM +/-0.02
|
|
Pressure
|
M2001 works with all Fluke 700 Series
Pressure Modules from 0-1" H2O to 10,000 psi
|
Module
Dependant
|
0.025% F.S. Typical
|
* All W /RTD inputs are 4W
**All Accuracy specs are "% of reading"
unless otherwise noted.
Fluke 700P Pressure Modules

Features:
• Display pressure readings in any of 10 different pressure units you specify in the calibrator setup
• Rugged urethane molded case protects the module from rough handling and harsh conditions
• Features internal temperature compensation from 0°C to 50°C for full-accuracy performance
• Includes NIST-traceable calibration certificate
• Module can be calibrated locally using the Fluke 700PCK Pressure Calibration Kit,
a PC, and a precision pressure source
• One year warranty

Order Modules By Individual Part Number Located in the First Column of the Above Table
|
OUR PRICE: THIS MODEL WAS DISCONTINUED REPLACED BY T-970
Ametek T-740 Pneumatic Calibration Pump Generates 200 PSIG
|
OUR PRICE: THIS MODEL WAS DISCONTINUED REPLACED BY T-975
Ametek T-750 Pneumatic Vacuum Calibration Pump Generates 0-25" Hg Vacuum
|
OUR PRICE: THIS ITEM WAS DISCONTINUED - WAS REPLACED BY T-975 (SEE BELOW)
Ametek T-930 Combination Pressure/Vacuum Calibration Pump generates 500 PSIG/28" Hg Vacuum
|
|
|
|
|
The Martel Beta Port-P Pressure Modules offer 29 standard pressure ranges from inches of water to 10000 PSIG, covering gauge, vacuum, absolute, compound, and differential measurements. Compatible with most Martel Beta Pressure calibrators
| |
|
|
|