Martel PTC-8001Thermocouple and RTD Calibrator

OUR PRICE: Replaced by Martel PTC-8010
Item Number: PTC-8001
Manufacturer: Martel Electronics (Beta)
Manufacturer Part No: PTC-8001

Martel PTC-8001 Thermocouple and RTD Calibrator

THE MARTEL PTC-8001 HAS BEEN REPLACED BY THE MARTEL PTC-8010

    The Martel PTC-8001 Precision Temperature Calibrator is an ideal device to handle all of your temperature calibration needs in a truely rugged, low cost package. It combines virtually all widely used thermocouples and RTD's in one device-you'll never have to worry about finding a temperature device you can't calibrate. This is especially true when calibrating "smart" or pulsed Current RTD transmitters where many other calibrators fail to work or operate at reduced accuracy specs. The PTC-8001 handles those applications with ease. 

Features: 

  • Thermocouple & RTD Source/Measure in One Unit
  • Works with "Smart" Pulsed Current Transmitters
  • High Accuracy +/-0.4 C J T/C and +/-0.3 C 4W Pt385 RTD 
  • 10 T/C Types and 8 RTD Types Including PT 385 (1000 ohm) and Cu (10 ohm)
  • Direct Keyboard Entry or Scroll of Output
  • NEMA 4 Rated Case, Rugged Design
  • YSI 400 Series Thermistor Support
  • Standard Mini-Plug or Banana Jack Connections for T/C
  • Store up to Nine Setpoints for Each Output Function
  • RS232 Interface for Remote PC Control
    Technical Specifications 

T/C Ranges J, K, E, R, S, N, B, L, U including - 10 to 70mV range
RTD Ranges Pt 385 (100, 200, 500, 1000ohms)
Pt 392, JIS, Ni 120, Cu10, YSI 400
Ohms Ranges 0 to 400.00 and 400.0 to 4000.0
Accuracy T/C Type J: +/-.3 C +/-10uV (+/-0.6 C total error)
RTD PT100: +/-.1 C +/-.075W (+/-0.3 C total error)
Ohms (400): +/-0.1 Ohm
Ohms (4000):+/-1.0 Ohm
RTS IEX-range 0.01mA to 3mA
RTD Frequency
Response
10 mS - works with al pulsed XMTR's
Oper. Temp. -10 C to 50 C
Storage Temp. -20 C to 60 C
Power Supply 4 AA Alkaline Cells, 30 hours typical operating time
Size 7.5" x 4" 1.5"


Range (4 Wire) Min Max Accuracy
Ohms Read (Low) 0.00 400.00 0.1 Ohm
Ohms Read (High) 400.0 1500.0 0.5 Ohm

1500.0 4000.0 1.0 Ohm

Range Min Max Excitation
Current
Accuracy
Ohms Source (Low) 5.00
5.00
400.00
400.00
0.1 to 0.5mA
0.5 to 3mA
0.15 Ohm
0.1 Ohm
Ohms Source (High) 400.0
1500.0
1500.0
4000.0
0.05 to 0.8mA
0.05 to 0.4mA
0.5 Ohm
1.0 Ohm


RTD Read and Source
Range Min Max Accuracy
Ni120 (672) -80.0 260.0 0.2 C
Pt100 (385) -200.0 800.0 0.33 C
Pt100(3926) -200.0 630.0 0.3 C
Pt100(3916) -200.0 630.0 0.3 C
Pt200(385) -200.0 630.0 0.8 C
Pt500(385) -200.0
500.0
500.0
630.0
0.3 C
0.4 C
Pt1000(385) -200.0
100.0
100.0
630.0
0.2 C
0.3 C
Cu10 -100.0 250.0 2.2 C
YSI400 15.00 50.00 0.05 C


Thermocouple Read and Source(All Errors Included)
Range Min Max Accuracy
J Thermocouple -200.0
0.0
0.0
1200.0
0.6 C
0.4 C
K Thermocouple -200.0
0.0
0.0
1370.0
0.8 C
0.5 C
T Thermocouple -200.0
0.0
0.0
400.0
0.8 C
0.5 C
E Thermocouple -200.0
0.0
0.0
950.0
0.5 C
0.4 C
R Thermocouple -20
0
500
0.0
500
1750
2.4 C
1.7 C
1.3 C
S Thermocouple -20
0
500
0
500
1750
2.4 C
1.7 C
1.4 C
B Thermocouple 600
800
1000
800
1000
1800
2.1 C
1.7 C
1.3 C
L Thermocouple -200
0.0
0.0
900.0
0.45 C
0.4 C
U Thermocouple -200
0.0
0.0
400.0
0.7 C
0.45 C
N Thermocouple -200
0.0
0.0
1300.0
1.1 C
0.6 C
mV Read/Source -10.00 75.00 0.015% +/- 2