New!!! Fluke 726 Advanced Process Calibrator
|
Item Number: Fluke 726
Manufacturer: FLUKE
Manufacturer Part No: 2452080
|
Fluke 726 Process Calibrator
The 726 measures and sources almost all process parameters and can
calibrate almost anything. The 726 will also interpret results without the help
of a calculator and store measurement data for later analysis.
Features:
• More precise measurement and calibration source performance, accuracies of 0.01%.
• Transmitter error% calculation, interpret calibration results without a calculator
• Memory storage for up to 8 calibration results, return stored calibration data from the field for later analysis
• Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
• HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
• Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
• Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement.
• New voltage input protection design for improved reliability
• Two separate channels; measure, source and view process signals simultaneously
• Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
• Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
• Measure or *source pressure using any of 29 Fluke 700Pxx Pressure Modules
• Source mA with simultaneous pressure measurement to conduct valve and I/P tests
• Perform fast linearity tests with auto step and auto ramp features
• Power transmitters during test using 24 V loop supply and simultaneous mA measurement
• Store frequently used test setups for later use
• Three-year warranty
*Pressure pump required
|
Function Measure or Source
|
Range or Type
|
Resolution
|
Accuracy
|
|
Voltage
|
0 to 100 mV 0 to 20 V (source)
0 to 30 V (measure)
|
0.01 mV 0.01 V 0.01 V
|
0.01 % Rdg + 2 LSD
|
|
mA
|
0 to 24
|
0.001 mA
|
0.01 % Rdg + 2 LSD
|
|
mV (TC terminals)
|
10.00 mV to +75.00 mV
|
.01 mV
|
0.01 % of range + 1 LSD
|
|
Resistance
|
5 Ω to 4000 Ohm
|
≤.01 Ω to
0.1 Ohm
|
0.015 %
|
|
Frequency
|
2.0 to 1,000 CPM
1 to 1100 Hz
1.0 to 10.0 kHz
10.0 to 15.0 kHz
|
0.1 CPM 1 Hz 0.1 kHz 0.1 kHz
|
+/- 0.05 % +/- 0.05 % +/- 0.25 % +/- 0.5 %
|
|
Loop Supply
|
24 V DC
|
N/A
|
10 %
|
|
Thermocouples
|
J, K, T, E, L, N, U
|
0.1 ºC, 0.1 ºF
|
to 0.2 ºC
|
|
Thermocouples
|
B, R, S
|
1 ºC, 1 ºF
|
to 1.2 ºC
|
|
RTDs
|
Cu10, Ni120 (672) Pt 100, 200, 500, 1000 (385) Pt 100 (3916), Pt 100 (3926)
|
0.01 ºC
0.01 ºF
|
to 0.15 ºC
|
|
Simultaneous Function Capability
|
Channel A
|
Channel B
|
|
24.000 mA DC
|
M
|
M or S
|
|
24.000 mA DC with 24 V loop supply
|
M
|
|
|
100.00 mV DC
|
|
M or S
|
|
30.000 V DC measure
|
M
|
|
|
20.000 V DC measure, 10.000 VDC Source
|
|
M or S
|
|
15 to 3200 Ohms
|
|
M or S
|
|
Thermocouple J, K, T, E, R, S, B, M, L, U, N
|
|
M or S
|
|
RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200, 500, 1000 (385)
|
|
M or S
|
|
Pressure (requires Fluke 700PXX Modules)
|
M
|
M used as S
|
|
Frequency; Squarewave, 1 CPM to10 kHz; fixed amplitude 5 V pp
|
|
M or S
|
Includes: TL75 Test Leads, AC72 Test Clips, one pair of stackable test leads, product overview
manual (print) and user’s manual (CDROM) in 14 languages.
|
|
|
Hard Instrument Carry Case with Pluck Foam to fit your instruments. 100's of sizes & styles available. Call for Details
|
|
|
|
|
|
|
|
|
|
|
|
The Fluke PV350 Allows any Multimeter to Read Vacuum and Pressure
|
|
|
|
|
|
|