Martel TC-100 Thermocouple Calibrator
|
Item Number: TC-100
Manufacturer: Martel Electronics (Beta)
Manufacturer Part No: TC-100
Personalization
* denotes required field
NIST Traceable Calibration*
|
Martel TC-100 Thermocouple Calibrator
The Martel TC-100 Thermocouple Calibrator provides high accuracy source and measurement of ten common thermocouples, as well as mV. The TC-100's accuracy of +/-0.3 C for Type J T/Cs includes all errors, at resolutions of +/-0.01 C or F in measure mode and +/-0.1 C or F in source mode. Features including MIN/MAX recall in measure mode, three setpoints per thermocouple range, a large knob for decade control of the output in source mode, and the ability toaccept bare T/C wires in addition to mini-plug inputs, makes the TC-100 an accurate, easy-to-use instrument for all your thermocouple calibration needs.

Features:
- High Accuracy - +/-0.3 C (Type J T/C all Errors Combined)
- Ten (10) Common T/C Types Plus mV
- Accepts Both T/C Mini-Plug and Bare T/C Wires
- Simple Decade Control of Output
- Three (3) Setpoints for each T/C Type
- MIN/MAX Recall in Measure Mode
- Input Protection to 240 VAC
- Passworded User Interface Allows Selction of Only Certain Functions
- Two Digit Resolution in measure mode ie 100.00
- Doubles as a Highly Accurate Thermometer
Technical Specifications
(1 year at 23 C +/-5 C; % of reading unless otherwise noted):
| Output Voltage |
Range: -10 to +75.000 mV
Resolution: 1 uV
Accuracy: +/-0.007 % of reading, +/-10 uV
Output Impedance: =1 Ohm |
| Input Voltage |
Range: -10 to +75.000 mV
Resolution: 1 uV
Accuracy: +/-0.007 % of reading, +/-10 uV
Input Impedance: > 1 MegOhm |
| Types: |
J, K, T, E, R, S, B, L, U, C |
| Range: |
mV |
| Resolution: |
Source +/-0.1 C or F
Measure +/-0.01 C or F |
| Accuracy: |
J +/-0.5 C; -210 C to -100 C
+/-0.3 C; -100 C to +1,200 C
K +/-0.6 C; -200 C to -100 C
+/-0.35 C; -100 C to +1,000 C
+/-0.5 C; +1,000 C to +1,372 C
T +/-0.7 C; -200 C to -150 C
+/-0.3 C; -150 C to +400 C
E +/-0.5 C; -200 C to -100 C
+/-0.3 C; -100 C to +1,000 C
R +/-1.8 C; 0 C to 250 C
+/-1.0 C; 250 C to +1,767 C
S +/-1.8 C; 0 C to 250 C
+/-1.0 C; 250 C to +1,767 C
B +/-1.7 C; 600 C to 1,000 C
+/-1.2 C; 1,000 C to 1,820 C
L +/-0.5 C; -200 C to -100 C
+/-0.4 C; -100 C to +900 C
U +/-0.7 C; -200 C to 0 C
+/-0.3 C; 0 C to +600 C
C +/-0.4 C; 0 to C 1,000 C
+/-0.7 C; 1,000 C to +1,800 C
+/-1.2 C; +1,800 C to +2,316 C |
| CJC Temperature Offset: |
+/-0.05 C/ C outside of 23+/-5 C |
| Warm-up Time: |
1 minute to specification |
| Environmental: |
Operating -10 C to +55 C
Storage -20 C to +70 C |
| Power Requirements: |
Battery 9 Volt alkaline; 006P/ IEC 6F22/
NEDA1604
Optional NiCad
Optional AC adapter/charger |
| Mechanical Dimensions: |
5.7" x 3.15" x 1.43"
(14.47 cm x 8.00 cm x 3.63 cm) |
| Weight: |
12 ounces (340 grams) |
| Notes: |
Temperature Standard ITS-90 |
|
Hard Instrument Carry Case with Pluck Foam to fit your instruments. 100's of sizes & styles available. Call for Details
|
|
|
|
|
|
|
|